NIIN Output Data
NIIN:
NSN:
Item Name:
VALVE , DEMAND RESUSCITATION
Definition:
AN ITEM DESIGNED FOR PROVIDING OXYGEN TO A BREATHING OR NON-BREATHING PATIENT IN AN EMERGENCY SITUATION. IT MAY INCLUDE A STANDARD INSIDE/OUTSIDE DIAMETER OUTLET FOR CONNECTION TO A MASK, TUBE, TRACHEAL OR A TUBE, TRACHEOSTOMY. IT IS COMPATIBLE WITH PORTABLE OXYGEN CYLINDERS WITH A REGULATOR, OR CENTRALIZED OXYGEN SOURCE. EXCLUDES VALVE, RESUSCITATOR.
Reference/Part Number
Part Number | CAGE | Status | RNCC | RNVC | DAC | RNAAC | MSDS | SADC |
---|---|---|---|---|---|---|---|---|
L034-100 | 62842 | A | 3 | 2 | 4 | 9Z |
CAGE Information
Code | Company |
---|---|
62842 | ALLIED HEALTHCARE PRODUCTS INC. |
Characteristics (Decoded)
MRC | Requirements Statement | Clear Text Reply |
---|---|---|
FEAT | SPECIAL FEATURES | DEMAND MODE 0 TO 160 AT 40 PSI; RESUSCITATION MODE 160 LPM AT 40 PSI; OPERATION TEMP MINUS 30 TO 125 DEG F; INLET FITTING 0.125 IN NPT FEMALE; FILTER 2 MICRON SINTERED STAINLESS STEEL; OUTLET 22 MM O/D BY 15 MM I/D; DELIVERPRESSURE 55 TO 65 CM/WATER; SUPPLY PRESSURE 50 PSI |
MATL | MATERIAL | PLASTIC, POLYSULFONE |
CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | LSPELDER CPR/DEMAND VALVE |
ANGD | DISPOSITION AFTER INITIAL USE | REUSABLE |